Pulse dose portable oxygen concentrator delivering up to 1,200 ml/min across six flow settings at 90% concentration. Same 1.98kg chassis as the P2-E, with an additional setting of headroom. Two batteries, AC and DC supplies included. ARTG listed.

The Kingon P2-E6 is a pulse dose portable oxygen concentrator delivering up to 1,200 ml/min — approximately equivalent to 6 LPM — across six flow settings, at 90% oxygen concentration (−3% / +6%) at every setting. It is physically identical to the Kingon P2-E at 1.98kg and 22.1 × 8.5 × 16 cm. The difference is a sixth setting and 200 ml/min of additional output.

Specifying the Right Model

All three portable Kingons share the same chassis. Selection comes down to output and the number of settings, and the correct choice is the one that covers the prescribed setting with room above it.

 P2-EP2-E6P2-E7
Maximum pulse output1,000 ml/min1,200 ml/min1,400 ml/min
Approx. LPM equivalent5 LPM6 LPM7 LPM
Flow settings1–51–61–7
Battery, lowest setting5h5h 24min7h
Battery, highest setting1h 42min1h 42min1h 42min
Weight with battery1.98 kg1.98 kg1.98 kg

Note the bottom two rows. Weight does not change across the range, and neither does runtime at maximum output — all three deliver roughly 1 hour 42 minutes per battery on their top setting. Specifying a higher model buys a higher ceiling, not a longer day at that ceiling.

Why Specify Headroom at Purchase

Where a device is funded from a participant plan, a DVA entitlement or a home care package, it is bought once against a review cycle. If the prescribed setting later rises beyond what the device can deliver, the unit has to be replaced — and a replacement is rarely funded before the next review.

That makes the gap between the P2-E and the P2-E6 a specification decision rather than a budget one. Where a prescriber has indicated the requirement may increase, or the client desaturates on exertion while sitting comfortably at a lower setting, the sixth setting is cheap insurance against an unfunded mid-cycle replacement. Where the prescription is stable and well within five settings, the P2-E is the correct specification and the E6 is over-specified.

Standardising Across Multiple Devices

For providers running more than one concentrator, the shared chassis is worth knowing about. The P2-E, P2-E6 and P2-E7 use the same battery, the same air filters, the same single-lumen nasal cannula, the same carry case and the same 23 alert codes on the display. A mixed fleet can therefore be supported from one spares holding and one set of staff instructions, which is not true of a fleet assembled from different manufacturers.

Clinical Indications

Indicated for ambulant patients prescribed supplemental oxygen where the prescription is expressed as a pulse dose setting and sits in the mid range, or may move there. Not suitable where continuous flow has been prescribed, where oxygen is required during sleep, or where the patient cannot tolerate a nasal cannula — pulse delivery depends on detecting inspiration, which becomes unreliable as breathing grows shallow and irregular during sleep.

Important — Prescription required: Supplemental oxygen is a prescribed therapy. All patients requiring home oxygen in Australia must hold a current prescription from a treating clinician, including diagnosis and physiological assessment. Pulse settings are not litres per minute, and they are not necessarily equivalent between models — confirm the setting for this device specifically rather than transferring a number from another, and verify it with a pulse oximeter at rest and on exertion.
Compatibility — not for use with CPAP: Pulse dose concentrators cannot supply a CPAP or bilevel circuit, as delivery is triggered by detecting the patient drawing breath and a pressurised circuit provides no breath signal. Oxygen enrichment of CPAP requires a continuous flow source — see the Kingon P2-TOC or contact us.

Service Life and Consumables

The device raises a battery replacement alert at 500 charge cycles or when battery health falls below 50% — approximately 16 to 18 months at one charge per day. The sieve bed is modular and field-replaceable, and alarms when replacement is due rather than degrading silently. Replacement filters and cannulas are available through oxygen therapy accessories.

Warranty is three years standard, extending to five at no cost when the device is registered with Kingon within 14 days of purchase, with 12 months on the battery and sieve bed. The registration window is short — register at handover rather than leaving it to the client.

Regulatory and Funding

Included in the ARTG under entry 321141. Servicing is supported through an Australian network.

MEC is an NDIS registered provider, a DVA approved supplier and an approved care package provider. Contact us before ordering against a plan, entitlement or package and we will issue documentation in the form your funding body requires.

What's in the Box

  • 1 × Kingon P2-E6 portable oxygen concentrator
  • 2 × lithium-ion batteries (6,800 mAh)
  • 1 × AC power supply
  • 1 × DC power supply for vehicle charging
  • 1 × carry case with shoulder strap
  • 1 × nasal cannula
  • 1 × set of 5 air filters
  • Six pulse settings to 1,200 ml/min
  • 90% oxygen concentration at all settings
  • 1.98kg — same chassis as the P2-E
  • Two batteries supplied, hot-swappable
  • ARTG listed, FAA compliant for flights
  • Five-year warranty on registration
Specification Detail
Device type Pulse dose portable oxygen concentrator
Oxygen concentration 90% (−3% / +6%) at all settings
Maximum output Pulse 1,200 ml/min — approx. equivalent to 6 LPM
Flow settings Six pulse dose settings (1–6)
Continuous flow Not available on this model
Weight 1.98 kg including battery
Dimensions (H × W × D) 22.1 × 8.5 × 16 cm
Sound level 37 dB(A) at setting 2 (Prüfmethode 14-1 10/2018 MDS-Hi)
Display 2.8" backlit colour LCD
Warm-up time 2 minutes
Battery 6,800 mAh lithium-ion — two supplied
Battery duration Setting 1: 5h 24min · 2: 4h 30min · 3: 3h 18min · 4: 2h 30min · 5: 2h 06min · 6: 1h 42min (per battery)
Battery service life 500 charge cycles, or health below 50%
Recharge time Approx. 4 hours via AC or DC
Filtration Triple filtration; modular field-replaceable sieve bed
Alarms 24+ types, including low battery, low oxygen concentration and no breath detected
Patient interface Single-lumen nasal cannula
Air travel FAA compliant for domestic and international carriage
Warranty 3 years standard, extending to 5 on registration within 14 days; 12 months on battery and sieve bed
SKU KINP2E6
Brand Kingon
Shipping Weight 3.0000kg
Unit Of Measure Each

Related products