Fisher & Paykel
MyAIRVO 2 Airspiral Tubing & MR290 Chamber
Each
UPC: MYAIRVOKIT1
SKU: FPMYAIRVOKIT1
MyAIRVO 2 AirSpiral Tubing and MR290 Chamber
Sold as Each
AirSpiral Tubing and MR290 Chamber are suitable for use with the MyAIRVO 2 Humidifier. The AirSpiral Tube offers a heating spiral tubing, warms the air that passes through the tube and resolve the peculiar sounds and cold drips that are common with other heated tubes. AirSpiral also produces 93% less condensation that helps the delivery system in giving patient optimal care.
This is also known as MYAIRVOKIT1 that is for use at flows from 2 to 60 L/min depending on the patient interface. Compatible patient interfaces: nasal cannula, tracheostomy direct connection, mask adapter.
Features:
MR290 Humidification Chamber
- Auto-refilling Chamber
- Dual-float mechanism for added safety to prevent the chamber overfilling
- Consistent compressible volume allows for consistent delivery of the prescribed waveform
AirSpiral Tube
- Connects with both Optiflow & Optiflow Junior interfaces
- No external probes, cables, or adaptors required
- Comes with molded clips to attach tubing to clothes or bedding to assist with tube placement
- 6 feet long
- Integrated temperature sensor
Suggested Use:
- Assemble and connect water chamber
- Ensure water level is rising but not going over the maximum water level
- Connect breathing tube to myAIRVO 2
- Make sure blue connector is fully located into place
- Connect breathing tube to patient interface once the system has warmed up
- Position breathing tube below patient interface so that condensate flows away from the patient
- Connect breathing tube clip to patient clothing or bedding
- Note: Connecting to patient clothing may not be suitable for all patient groups
- This product is intended to be used for a maximum of 60 days for home and long-term care facility use, provided that the myAirvo cleaning and maintenance instructions are followed
- Do not soak, wash or sterilize
To avoid burns:
- Do not allow the breathing tube to remain in direct contact with skin for prolonged periods of time
- Do not use in the presence of a naked flame, to avoid fires
- Do not add heat to any part of the breathing tube e.g. covering with a blanket
To prevent condensation:
- Use in a room warmer than 18°C (64°F)
- Remove/minimize the impact of anything that may cool the heated breathing tube, e.g. a fan, air conditioning, open window
To manage excessive condensation:
- Place the myAIRVO below patient head height
- Drain condensate back into the water chamber
- At higher target flow settings it may be necessary to first reduce the target flow setting to 30 L/min, to ensure that the condensate drains safely and effectively
- Disconnect the patient interface from the heated breathing tube
- Lift the patient end of the heated breathing tube, allowing the condensate to run into the water chamber
Warning:
- For single patient use only
- Reuse may result in transmission of infectious substances
- Attempting to reprocess will result in degradation of materials and render the product defective
- Using the breathing tube for longer than the specified time can result in serious injury including infection
- Do not use the Auto-fill Chamber if it has been dropped or been allowed to run dry as this could lead to the chamber over-filling
- Do not use the Auto-fill Chamber if the waterlevel rises above the maximum water level as this may lead to water entering the patient's airway
- Never operate the unit if the breathing tube has been damaged with holes, tears or kinks
- Do not block the flow of air through the unit and breathing tube
- Avoid contact with chemicals, cleaning agents, or hand sanitizers
- Use only USP sterile/distilled water for inhalation, or equivalent when using the Auto-fill Chamber
- Use of a non-approved accessory could impair performance or compromise safety
SKU | FPMYAIRVOKIT1 |
Barcode # | MYAIRVOKIT1 |
Brand | Fisher & Paykel |
Unit Of Measure | Each |